A9 Instrument Error Setting Information Prompt

1.Err 13 indicates that the division value is set incorrectly. Please reset the division value. 2.Err 14 means: The decimal place setting must be less than 5. Please reset the decimal place. 3. Err 15 indicates that the overload alarm value should not be less than 100. Please reset the overload alarm value. 4.Err 17 means: a. The overload alarm value should not exceed 325000. Please reset the overload alarm value. b. The input value is greater than the allowed value, please re-enter the data.

A9 Other information prompts on the instrument

1.Err 24 indicates that the calibration head should not be inserted when the instrument is working normally. If calibration is required, the calibration head should be inserted after powering on and removed before shutting down. When the instrument is turned on and the calibration head is inserted, it displays Err 24 for about 6 seconds to remind the operator to pay attention. 2.Err 25 indicates that illegally copied software or E2PROM is damaged. 3. ctnu 0 means: In the 8th or 9th step of calibration, if the instrument still cannot collect data that it considers stable within 25 seconds, this display will be made. At this point, the operator can input 0, 1, or 2, which have the following functions: Enter 0: (Abort) to tell the instrument that there is no need to do this step again and proceed to the next step. Enter 1: (Repeat) to tell the instrument panel to try again. Enter 2: (Lgnore) tells the instrument that these unstable data can be used.

A9 Component Malfunctions and Troubleshooting Information Tips

1.Err 18 indicates that the keyboard is faulty and will enter the weighing display after 10 seconds. The keyboard needs to be replaced. 2.Err 20 indicates that some data in RAM is lost. After turning the calibration switch to the calibration state and restarting the self-test, Err 20 will no longer appear. Then, turn the calibration switch to the non calibration state. 3. Err 21 indicates that there is a loss of calibration data for both RAM and E2PROM. In this case, the calibration plug must be plugged in, the original calibration data must be re entered, and then the machine can be turned on or recalibrated. 4.Err 22 indicates that the RAM is damaged and a new chip must be replaced before recalibration. 5.Err 23 indicates that the E2PROM is damaged and a new chip must be replaced to re calibrate the original data After input, restart or recalibrate. 6. Err SP indicates that there is a malfunction with the built-in micro printer, press any key to exit. Please check the internal printing connection or Replace the printer head and driver board.

A9 incorrect connection message prompt

1.Err P means: Printer connection error or printer error, press any key to exit, reconnect or replace the printer. 2.Err 01 indicates that the sensor signal is wired incorrectly or the sensor signal is biased negative beyond the input signal range of the instrument (1) If this is a scale in use, it can be determined that there is a wiring fault or a damaged sensor. (2) If this is an untested scale, please carefully check if the sensor wiring is correct and if the sensor is working properly. 3. Err 02 indicates that the sensor signal connection is incorrect or the signal exceeds the A/D range: (1) If this is a scale in use, it can be determined that there is a wiring fault or a damaged sensor. (2) If this is an untested scale, you can check it in the following order: a. Carefully check if the sensor wiring is correct. b. Whether the selection of sensors is reasonable should meet the requirement that the empty load of the sensor+the amount of the scale